NABEL-032H1RN1TZ NABEL-032H1RN1Z NABEL-092H1RN1GT NABEL-052H1RN1G NABEL-242H1RN1 NABEL-152H1RN1GT NABEL-032H1RN1G NABEL-402H1RN1T NABEL-092H1RN1G NABEL-242H1RN1TZ NABEL-092H1RN1 NABEL-242H1RN1GT NABEL-032H1RN1GT 455-0653 NABEL-052H1RN1Z NABEL-242H1RN1T NABEL-402H1RN1TZ NABEL-152H1RN1Z NABEL-152H1RN1G NABEL-242H1RN1Z NABEL-052H1RN1TZ NABEL-402H1RN1G 455-0655 NABEL-402H1RN1Z NABEL-032H1RN1 NABEL-152H1RN1TZ 455-0651 NABEL-402H1RN1 NABEL-052H1RN1 NABEL-092H1RN1Z NABEL-402H1RN1GT 455-0649 NABEL-092H1RN1TZ NABEL-052H1RN1GT NABEL-152H1RN1 NABEL-242H1RN1G
Muffle furnaces with lift up door, Tmax 1100 °C and standard controller B410, LT 3/11 - LT 40/11
Piece i suszarki Piece muflowe i piece do spopielania
The LT 3/11 - LT 40/11 muffle furnaces are the ideal choice for daily laboratory use. These high quality models have an advanced and attractive design, and are very reliable. The muffle furnaces have a lift up door as standard.

  • Textured stainless steel housing
  • Dual shell housing for low external temperatures and high stability
  • Lift door (LT models) with hot surface facing away from the operator
  • Exhaust air outlet on the rear wall of furnace
  • Solid state relays provide low-noise operation
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: Recording of process data with USB-flash drive

Heating is from two sides by means of ceramic heating plates (heating from three sides for muffle furnaces LT 24/11 - LT 40/11). These ceramic heating plates with an integral heating element are protected from fumes and splashing, and are easy to replace. Only fibre materials which are not classified as carcinogenic according to TRGS 905, class 1 or 2 are used.

The furnace is equipped with a B410 standard controller with LC display, featuring 5 freely adjustable programs and 4 segments and 2 extra functions. These controllers are PID microprocessor controlled with a self-diagnostic system and integrated process documentation. The B410 controllers can automatically activate the vent with fan or catalyst (not model LT 15).

Informacje do zamówienia: All units require a 230 V single-phase power supply, except LT 24 and LT 40 (400 V 3/N/PE). Optional accessories e.g. a vent, vent with fan or catalytic converter, over-temperature limit controller with adjustable temperature cut out for thermal protection Class 2, in accordance with EN 60519-2 as temperature limiter to protect the oven and load, protective gas connection on the rear wall of furnace. Manual or automatic gas supply systems are also available.

Uwaga: Maximum continuous operating temperature should be 50 °C below the 'Max. temp.' of the model.
Note heat up times: If connected at 230 V 1/N/PE rsp. 400 V 3/N/PE.
Outer dimensions include opened lift up door.